400 008 3510
如果您有任何疑问请及时与我们联系!产品优势: Taiwanese craftsmanship, imported accessories, small size, light weight, convenient movement, fast response, high accuracy, and extremely high stability
生产能力: 100 units/month
应用领域: Suitable for use with medium-sized three-phase medical equipment such as DR, X-ray machine, CR, patient monitoring equipment, PET, digital gastrointestinal machine, etc.
Sohu Net: DR, CT, MRI equipment specific voltage regulators
Sohu Net: The Importance of Regulators and UPS Power Supplies for Medical Devices
NetEase: Why medical devices need to be equipped with dedicated voltage regulators
Jingyuan Technology has always been paying attention to the serious impact of China's power grid on medical equipment. As medical equipment is a precision instrument used for detecting, diagnosing, and treating the human body, its stability and accuracy can cause health and even life losses to humans if it affects the equipment. So we have customized personalized specialized voltage regulators for medical industry users, using more stable power supply to achieve mutual benefit and increase the value of their assets. It has the characteristics of phase separation regulation, high voltage regulation accuracy, true effective value sampling, and strong instantaneous load impact resistance.
Sohu Net: DR, CT, MRI equipment specific voltage regulators
Sohu Net: The Importance of Regulators and UPS Power Supplies for Medical Devices
NetEase: Why medical devices need to be equipped with dedicated voltage regulators
Models: JBW-50KVA, JBW-80KVA
Capacity: 50KVA, 80KVA
Input voltage: 380V ± 15% (above 15%, please specify in advance)
Output voltage: 380V (1-5% adjustable)
Output accuracy: ± 1.5%
Rated current: 76A, 121A
Display: Output voltage display, overvoltage/working/undervoltage status display
Efficiency:>98%
Frequency: 50Hz/60Hz
Load power factor: 0.8
Reaction time: < 0.04 seconds
Adjustment time: > 20V/S
Waveform distortion: < 0.1%
Equipment noise: ≤ 45DB
Voltage adjustment: three-phase split adjustment, can automatically adjust the three-phase balanced output when the input three-phase is unbalanced
Craft type: Taiwanese craft
Temperature rise: < 75 ℃
Cooling method: forced air cooling
Insulation resistance: 5M Ω
Insulation level: Class B
Dielectric strength: AC2000V/1min
Protection: overcurrent/short circuit protection, overvoltage/undervoltage alarm
model | capacity | dimensionsW*L*H(mm) | weight(kg) |
---|---|---|---|
JBW-50KVA | 50KVA | 450*750*860 | 280 |
JBW-80KVA | 80KVA | 500*750*900 | 300 |
The above data is for reference only, and I will make adjustments according to the actual situation.
1. Environmental temperature: 0 ℃~40 ℃
2. Humidity: 0-90%
3. Altitude: not exceeding 2000m
4. Installation site: Gas, steam, dust, dirt, and other explosive and corrosive media without severe insulation
5. Installation site: There should be no severe vibration or bumps
6. Other instructions: Any special usage conditions that do not meet the above regulations should be determined through consultation between the user unit and our factory